Universal Temperature Regulation Cryogenic System “UTRECS” for microoptical measurements

|
“UTRECS” A-287
“UTRECS” A-287 is used with the optical microscope Olympus BX41 for microoptical measurements in a temperature range between 5 K and 350 K. “UTRECS” A-287 is a patent pending system in Russia and Ukraine.
|
Features:
|
|
Temperature range |
8 – 350 К (дискретність: < 0,1 К) |
Temperature stability |
±0,01 K |
Diameter of sample holder |
15 mm |
Cryoagents |
liquid He (8 – 350 K) liquid N2 (80 – 350 K) |
He consumption (10 K) |
< 0.1 liter/hour |
N2 (80 K) |
< 0,2 liter/hour |
Overall dimensions |
590 х 313 х 160 mm |
Net weight |
< 8 kg
|
|
Applications:
- material science;
- solid-state physics;
- biophysics and medicine.

|
“UTRECS” C-40
“UTRECS” C-40 is a continuous-flow He-N2 cryogenic system used with optical microscopes for microoptical measurements in a temperature range between 40 K and 1000 K.
|
Features:
|
|
Temperature range |
40 – 1000 К (resolution : < 1K) |
Temperature stability |
±0,01 K |
Diameter of sample holder |
15 mm |
Cryoagents |
liquid He (40 – 350 K) liquid N2 (80 – 350 K) without cryoagent (350 – 1000 K) |
Dimensions |
diameter:< 150 mm; height: < 70 mm |
Mass |
< 10 kg |
|
Applications:
- material science;
- solid-state physics;
- biophysics;
- pharmaceutics.

|
“UTREKS” B-30
“UTREKS” B-30 is used with the scanning electron microscope REMMA-102 for microoptical measurements in a temperature range between 5 K and 300 K. “UTREKS” B-30 is a patent pending system in Ukraine.
|
Features:
|
|
Temperature range |
5 – 300 К (resolution: < 0,15К) |
Cool down time (5 К) |
< 1 hour |
Temperature stability |
< 0.1 K |
Diameter of sample holder |
15 mm |
X-Y shift |
±7 mm |
Dimensions |
565 x 92 mm |
Net weight |
< 1,5 kg |
|
Applications:
- material science;
- solid-state physics.
Developer:
Cryogenic Technologies Laboratory of the Institute of Physics (I.P. Zharkov).
Contact information:
Tel. +380 44 525-98-41
E-mail: fesenko@iop.kiev.ua